AN132
Using the Network Analyzer as a Grid Dip Oscillator
Written by Bill Ashley
antenna analyzer, network analyzer  Save Application Notes (PDF)

Abstract:

This paper explains how to use a network analyzer (NA) as a grid dip oscillator (GDO).

Introduction:

The GDO originally used a triode vacuum tube configured as an oscillator, with a sensitive DC current meter measuring its grid current. Most versions found today are solid state (JFET) and have no grid (or grid current), but the name has stuck. This schematic shows an example of a GDO.

In this circuit, L1 resonates with C1, creating an oscillation. In operation, the RF energy in L1 couples into a tuned circuit under test. When the circuit under test absorbs energy from L1, the gate current in Q1 drops, causing the needle on M1 to dip. L1 plugs into the circuit because GDOs come with a set of inductors to cover several bands of frequencies.


Discussion:

A network analyzer normally has neither a meter nor an external coil for an oscillator. To get the network analyzer to operate like a GDO, we try to couple the S11 port of the network analyzer to the circuit under test and look for a shift in the impedance level. As with any GDO, there must be a strong coupling between the instrument and the load to get a noticeable “dip”. Use a wire loop probe mounted to the end of a coaxial cable to couple the network analyzer to the circuit under test. The exact design of the loop depends on the frequency range of interest. Try to make the loop with a Z of 30 to 70 ohms for best results. Since this application requires strong coupling, the shape factor of the loop depends on physical properties of the circuit under test.

Network Analyzer with GDO Sensing Loop

Resistance and Reactance of Sensing Loop , Uncoupled

Resistance and Reactance of Sensing Loop , Coupled into Resonant Circuit

Conclusion:

A network analyzer may be used in a manner similar to a GDO.

 

Copyright © 2001-2012 AEA Technology, Inc.
ALL RIGHTS RESERVED


antenna analyzer, network analyzer Antenna Testing
antenna analyzer, network analyzer Antenna Isolation Testing
antenna analyzer, network analyzer Cable Fault Locator
antenna analyzer, network analyzer Cable Tester
antenna analyzer, network analyzer Cable Measuring
antenna analyzer, network analyzer Circuit Analysis
antenna analyzer, network analyzer Circuit Gain/Loss
antenna analyzer, network analyzer MRI Coil Testing
antenna analyzer, network analyzer Network Analyzer
antenna analyzer, network analyzer Power Meter Reading
antenna analyzer, network analyzer Spectrum Analysis
antenna analyzer, network analyzer Vector Network Analyzer
antenna analyzer, network analyzer NMR Coil Testing/Tuning
antenna analyzer, network analyzer Amateur
antenna analyzer, network analyzer Aviation
antenna analyzer, network analyzer CATV/Broadband
antenna analyzer, network analyzer Government
antenna analyzer, network analyzer Laboratory
antenna analyzer, network analyzer Military
antenna analyzer, network analyzer MRI or NMR Coil Testing
antenna analyzer, network analyzer Railroad
antenna analyzer, network analyzer RFID
antenna analyzer, network analyzer Telco
antenna analyzer, network analyzer Wireless (Broadcast)
antenna analyzer, network analyzer Wireless (Land Mobile)
Access and use of this site is subject to the terms and conditions of our PRIVACY POLICY
Property of and Proprietary to AEA Technology, Inc.
Copyright © 2001-2012 AEA Technology, Inc.
ALL RIGHTS RESERVED
Privacy Policy Terms of use